Scientific Background
CD14 exists in two forms. It is either anchored into the membrane by a glycosylphosphatidylinositol tail (mCD14) or it appears in a soluble form (sCD14). Soluble CD14 appears either after shedding of mCD14 (48 KDa) or is directly secreted from intracellular vesicles (56 KDa).CD14 takes its name from its inclusion in the cluster of differentiation group of cell surface marker proteins. CD14 was the first described pattern recognition receptor.CD14 acts as a co-receptor (along with the Toll-like receptor TLR 4 and MD-2) for the detection of bacterial lipopolysaccharide (LPS). CD14 can only bind.
Assay Principle
This assay employs a two-site sandwich ELISA to quantitate CD14 in samples. An antibody specific for CD14 has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and anyCD14 present is bound by the immobilized antibody. After removing any unbound substances, a biotin-conjugated antibody specific for CD14 is added to the wells. After washing, Streptavidin conjugated Horseradish Peroxidase (HRP) is added to the wells. Following a wash to remove any unbound avidin-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of CD14 bound in the initial step. The color development is stopped and the intensity of the color is measured.
Performance Specifications
| Sensitivity | 1.52 ng/mL |
|---|---|
| Detection Range | 3.75-60 ng/mL |
| Total Assay Time | 3.5–5 hours |
| Compatible Sample Types | serum, plasma, cell culture supernatant, tissue homogenate |
| Species Reactivity | Human |
| Detection Method | Colorimetric (TMB/HRP) |
| Storage | 4°C (short-term); -20°C (long-term) |
✓ Research-Grade Validation
Safety & Regulatory
Handle reagents in accordance with institutional biosafety guidelines. Refer to the Safety Data Sheet (SDS) for complete hazard and handling information. Contains components that may require special disposal procedures per local regulations.
This kit is validated for use with serum, plasma, cell culture supernatant, tissue homogenate. For unlisted matrices (e.g., tissue lysate, urine), perform a spike-and-recovery experiment to confirm assay performance before generating reportable data. Sample dilution in the kit's provided diluent is recommended to minimize matrix interference.
The minimum detectable concentration (sensitivity) of this kit is 1.52 ng/mL. Values below this threshold should be reported as below the limit of detection (<LOD) and should not be extrapolated from the standard curve.
The total assay time from sample addition to absorbance reading is approximately 3.5–5 hours, including all incubation, wash, and substrate steps. Hands-on time is typically 1–2 hours; most steps involve passive plate incubation. Plan the assay as a single uninterrupted session for best results.
Standard components of this Sandwich ELISA Kit typically include: pre-coated microplate (96-well strip format), lyophilized or liquid recombinant CD14 standard, detection antibody, streptavidin-HRP conjugate, TMB substrate, stop solution, wash buffer concentrate, and sample/standard diluent. Refer to the kit insert or datasheet for the exact component list and storage requirements.
This kit uses colorimetric (TMB/HRP) detection and requires a standard microplate absorbance reader capable of measuring at 450 nm. A reference wavelength of 570 nm or 630 nm is recommended to reduce background. No specialized fluorescence or luminescence reader is needed. Ensure the instrument is calibrated and the plate is clean and free of condensation before reading.
Can’t Find What You’re Looking For? We can help you source the best match or customize an ELISA solution for your study. Options may include alternative target synonyms, different species reactivity, sample type/matrix compatibility (serum/plasma/lysate/supernatant), assay format (sandwich/competitive), sensitivity/range, detection chemistry (colorimetric/fluorescent/chemiluminescent), plate format (pre-coated/uncoated, strips vs full plate), and bulk or custom packaging. Click Talk to a Scientist to submit a request form, email us at support@biohippo.com, or explore our Research Services for additional support. Our team will be in contact with you shortly.