| Field | Specification |
|---|---|
| Mfr No | |
| Alternative Names | CC10; CC16; CCSP; UGB; Uteroglobin (Clara-cell specific 10-kD protein) |
| Product Type | |
| Sensitivity | |
| UniProt # |
Scientific Background
Clara Cell Protein 16 concentrations have been determined in both serum and bronchoalveolar lavage fluid in numerous studies since 1994. In serum, its increase is associated with age, asbestos, nitrogen chloride and ozone exposure, sarcoidosis and high PEEP ventilation. Decreased serum CC16 levels are found after pulmonary resection, in silica-exposed workers, smokers and in asthma. Decreased CC16 concentrations were also found in the amniotic fluid of fetuses suffering from pulmonary hypoplasia caused by various mechanisms (diaphragmatic hernia, diabetic fetopathy, Turner and Down syndrome).
Assay Principle
This assay employs a two-site sandwich ELISA to quantitate SCGB1A1 in samples. An antibody specific for SCGB1A1 has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and anySCGB1A1 present is bound by the immobilized antibody. After removing any unbound substances, a biotin-conjugated antibody specific for SCGB1A1 is added to the wells. After washing, Streptavidin conjugated Horseradish Peroxidase (HRP) is added to the wells. Following a wash to remove any unbound avidin-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of SCGB1A1 bound in the initial step. The color development is stopped and the intensity of the color is measured.
Performance Specifications
| Sensitivity | 0.067 ng/mL |
|---|---|
| Detection Range | 0.156-10 ng/mL |
| Total Assay Time | 3.5–5 hours |
| Compatible Sample Types | serum, plasma, cell culture supernatant, tissue homogenate |
| Species Reactivity | Human |
| Detection Method | Colorimetric (TMB/HRP) |
| Storage | 4°C (short-term); -20°C (long-term) |
✓ Research-Grade Validation
Safety & Regulatory
Handle reagents in accordance with institutional biosafety guidelines. Refer to the Safety Data Sheet (SDS) for complete hazard and handling information. Contains components that may require special disposal procedures per local regulations.
This kit is validated for use with serum, plasma, cell culture supernatant, tissue homogenate. For unlisted matrices (e.g., tissue lysate, urine), perform a spike-and-recovery experiment to confirm assay performance before generating reportable data. Sample dilution in the kit's provided diluent is recommended to minimize matrix interference.
The minimum detectable concentration (sensitivity) of this kit is 0.067 ng/mL. Values below this threshold should be reported as below the limit of detection (<LOD) and should not be extrapolated from the standard curve.
The total assay time from sample addition to absorbance reading is approximately 3.5–5 hours, including all incubation, wash, and substrate steps. Hands-on time is typically 1–2 hours; most steps involve passive plate incubation. Plan the assay as a single uninterrupted session for best results.
Standard components of this Sandwich ELISA Kit typically include: pre-coated microplate (96-well strip format), lyophilized or liquid recombinant SCGB1A1 standard, detection antibody, streptavidin-HRP conjugate, TMB substrate, stop solution, wash buffer concentrate, and sample/standard diluent. Refer to the kit insert or datasheet for the exact component list and storage requirements.
This kit uses colorimetric (TMB/HRP) detection and requires a standard microplate absorbance reader capable of measuring at 450 nm. A reference wavelength of 570 nm or 630 nm is recommended to reduce background. No specialized fluorescence or luminescence reader is needed. Ensure the instrument is calibrated and the plate is clean and free of condensation before reading.
Can’t Find What You’re Looking For? We can help you source the best match or customize an ELISA solution for your study. Options may include alternative target synonyms, different species reactivity, sample type/matrix compatibility (serum/plasma/lysate/supernatant), assay format (sandwich/competitive), sensitivity/range, detection chemistry (colorimetric/fluorescent/chemiluminescent), plate format (pre-coated/uncoated, strips vs full plate), and bulk or custom packaging. Click Talk to a Scientist to submit a request form, email us at support@biohippo.com, or explore our Research Services for additional support. Our team will be in contact with you shortly.