TCTP, is an essential component of the tuberous sclerosis-RHEB pathway. Reducing Drosophila Tctp levels reduced cell size, cell number, and organ size, which mimics Drosophila Rheb mutant phenotypes. Drosophila Tctp is genetically epistatic to Tsc1 and Rheb, but acts upstream of S6K, a downstream target of Rheb. Drosophila Tctp directly associated with Rheb and displayed guanine nucleotide exchange activity with it in vivo and in vitro. Human TCTP showed similar biochemical properties compared to Drosophila Tctp and could rescue Drosophila Tctp mutant phenotypes, suggesting that the function o.
Assay Principle
This assay employs a two-site sandwich ELISA to quantitate TPT1 in samples. An antibody specific for TPT1 has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and anyTPT1 present is bound by the immobilized antibody. After removing any unbound substances, a biotin-conjugated antibody specific for TPT1 is added to the wells. After washing, Streptavidin conjugated Horseradish Peroxidase (HRP) is added to the wells. Following a wash to remove any unbound avidin-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of TPT1 bound in the initial step. The color development is stopped and the intensity of the color is measured.
Performance Specifications
| Sensitivity | 0.113 ng/mL |
|---|---|
| Detection Range | 0.312-20 ng/mL |
| Total Assay Time | 3.5–5 hours |
| Compatible Sample Types | serum, plasma, cell culture supernatant, tissue homogenate |
| Species Reactivity | Mouse |
| Detection Method | Colorimetric (TMB/HRP) |
| Storage | 4°C (short-term); -20°C (long-term) |
✓ Research-Grade Validation
Safety & Regulatory
Handle reagents in accordance with institutional biosafety guidelines. Refer to the Safety Data Sheet (SDS) for complete hazard and handling information. Contains components that may require special disposal procedures per local regulations.
This kit is validated for use with serum, plasma, cell culture supernatant, tissue homogenate. For unlisted matrices (e.g., tissue lysate, urine), perform a spike-and-recovery experiment to confirm assay performance before generating reportable data. Sample dilution in the kit's provided diluent is recommended to minimize matrix interference.
The minimum detectable concentration (sensitivity) of this kit is 0.113 ng/mL. Values below this threshold should be reported as below the limit of detection (<LOD) and should not be extrapolated from the standard curve.
The total assay time from sample addition to absorbance reading is approximately 3.5–5 hours, including all incubation, wash, and substrate steps. Hands-on time is typically 1–2 hours; most steps involve passive plate incubation. Plan the assay as a single uninterrupted session for best results.
Standard components of this Sandwich ELISA Kit typically include: pre-coated microplate (96-well strip format), lyophilized or liquid recombinant TPT1 standard, detection antibody, streptavidin-HRP conjugate, TMB substrate, stop solution, wash buffer concentrate, and sample/standard diluent. Refer to the kit insert or datasheet for the exact component list and storage requirements.
This kit uses colorimetric (TMB/HRP) detection and requires a standard microplate absorbance reader capable of measuring at 450 nm. A reference wavelength of 570 nm or 630 nm is recommended to reduce background. No specialized fluorescence or luminescence reader is needed. Ensure the instrument is calibrated and the plate is clean and free of condensation before reading.
Can’t Find What You’re Looking For? We can help you source the best match or customize an ELISA solution for your study. Options may include alternative target synonyms, different species reactivity, sample type/matrix compatibility (serum/plasma/lysate/supernatant), assay format (sandwich/competitive), sensitivity/range, detection chemistry (colorimetric/fluorescent/chemiluminescent), plate format (pre-coated/uncoated, strips vs full plate), and bulk or custom packaging. Click Talk to a Scientist to submit a request form, email us at support@biohippo.com, or explore our Research Services for additional support. Our team will be in contact with you shortly.