| Field | Specification |
|---|---|
| Applications | |
| Storage | |
| Shelf life | |
| Shipping | |
| Catalog no. (Mfr.) | |
| Main SKU |
Product Overview
A telomere is a DNA sequence at the end of a eukaryotic chromosome composed of multiple repeating nucleotide elements (TTAGGG) in tandem. In addition to providing a buffer for non-transcribed DNA, it also protects chromosome ends from fusion and degradation, preserving chromosome stability and genetic integrity. Telomere length is determined initially by genetic and environmental factors and decreases over time. Studies have shown that telomere length is closely related to DNA repair, aging, apoptosis, and tumorigenesis. Therefore, accurate and repeatable measurement of telomere length is particularly important for researchers. This product mainly uses relative quantitative qPCR to directly compare the average telomere length of the sample, that is, the copy number ratio (T/S) of the telomere repeat sequence (Tel) and the genome single copy gene (SCR) copy number (T/S) is used as the telomere relative length. Single-copy gene primers (SCR) specifically recognize and amplify the 78 bp long region on human chromosome 11. The primer set in the kit has passed the test to ensure: (1) efficient and reliable quantification; (2) no non-specific amplification. Each set of primers has been verified by amplification curve efficiency (E>98%, R²>0.99), melting curve analysis and gel electrophoresis verification.
The assay is run in a 20 µL reaction with 0.5–5 ng of genomic DNA and a final primer concentration of 0.2 µM. Cycling is 95°C for 3 min, then 35–40 cycles of 95°C for 5 sec, 58°C for 30 sec and 72°C for 30 sec, followed by melting curve analysis. Relative telomere length is calculated by the comparative ∆∆Cq method as 2^(-∆∆Cq).
Specifications
| Method | Relative quantitative qPCR with melting curve analysis |
|---|---|
| Target | Telomere repeat sequence (Tel) against a genomic single-copy reference gene (SCR) |
| Species | Human |
| Reference gene region | 78 bp region on human chromosome 11 |
| Sample input | 0.5–5 ng genomic DNA per 20 µL reaction |
| Primer performance | E>98%, R²>0.99 |
| Result principle | Comparative ∆∆Cq; relative telomere length = 2^(-∆∆Cq) |
| ROX | 50× ROX supplied separately; added according to the instrument |
| Reactions per kit | 100 rxn or 500 rxn |
Kit Components
| Component | 100 rxn | 500 rxn |
|---|---|---|
| 2× Mix | 1 mL | 1.25 mL × 4 |
| Telomere Primer Mix (10 µM) | 50 µL | 100 µL |
| Single Copy Reference Primer Mix (10 µM) | 50 µL | 100 µL |
| 50× ROX | 250 µL | 1.25 mL |
| RNase-free ddH₂O | 1 mL | 1.25 mL × 4 |
Applications
- Relative telomere length measurement in human samples
- Comparison of average telomere length between samples
Storage & Handling
Stored at -20°C and protected from light, with a shelf life of one year. Shipping: Ice pack or dry ice.
Shipping Stability Validation
To support delivery to regions with long or unpredictable transit times, ELK Biotechnology ran a 15-day accelerated stability test on the 500-reaction kit (20 µL reactions). Kits were held either at 2–8 °C (refrigerated at 4 °C, simulating transport with intact ice packs) or at room temperature (simulating transport after ice-pack failure). At each time point, both assays used to calculate the T/S ratio, the Telomere (Tel) primer assay and the Single-Copy Reference (SCR) primer assay, were run and compared with a Day 0 baseline.
Maximum Ct shift vs. Day 0: |ΔCt| ≤ 0.61 for both assays under both storage conditions
2–8 °C storage (time points: 0, 6, 9, 12 and 15 days)
| Storage time | SCR mean Ct | SCR ΔCt | Tel mean Ct | Tel ΔCt |
|---|---|---|---|---|
| Day 0 | 25.91 | baseline | 21.83 | baseline |
| Day 6 | 26.23 | +0.32 | 21.22 | −0.61 |
| Day 9 | 26.36 | +0.45 | 21.34 | −0.49 |
| Day 12 | 26.17 | +0.26 | 21.34 | −0.49 |
| Day 15 | 26.31 | +0.40 | 21.62 | −0.21 |
Room-temperature storage (time points: 0, 3, 6, 9, 12 and 15 days, tested in two runs, each with its own Day 0 baseline)
| Run | Storage time | SCR mean Ct | SCR ΔCt | Tel mean Ct | Tel ΔCt |
|---|---|---|---|---|---|
| Run 1 | Day 0 | 25.81 | baseline | 21.94 | baseline |
| Run 1 | Day 3 | 25.98 | +0.17 | 22.06 | +0.12 |
| Run 1 | Day 6 | 26.16 | +0.35 | 21.83 | −0.11 |
| Run 2 | Day 0 | 26.03 | baseline | 20.50 | baseline |
| Run 2 | Day 9 | 26.29 | +0.26 | 20.05 | −0.45 |
| Run 2 | Day 12 | 26.50 | +0.47 | 20.27 | −0.23 |
| Run 2 | Day 15 | 26.62 | +0.59 | 20.66 | +0.16 |
Amplification was normal for both assays at every time point, and the manufacturer recorded every time point as a pass. The reagents stayed clear, with no precipitate, turbidity or discoloration, and amplification curves were smooth with a stable baseline throughout.
What this means for your order: these data support kit performance after transit delays of up to 15 days, including loss of the cold chain at room temperature. They do not change the recommended storage condition: on receipt, store the kit at -20°C, protected from light. Because relative telomere length is calculated against a reference sample, run all samples and the reference DNA of one study with the same kit lot, and on the same plate where possible.
Source: ELK Biotechnology stability validation report for catalog no. EQ022 (November 2026). ΔCt = mean Ct at the time point minus mean Ct at Day 0 of the same run. Values are mean Ct as reported by the manufacturer; the number of replicates, standard deviations and the exact room temperature were not stated in the report.
Usage Notes
- At least three biological replicates are required.
- Avoid strong light during storage and reaction preparation.
- Wear a lab coat and disposable operating gloves during the procedure.
- Confirm the amplification and melting curves after the reaction.
Safety & Regulatory
For research use only. Not intended for clinical, diagnostic or therapeutic use. Follow the manufacturer's manual and standard laboratory safety practice.
Customization & Add-ons: Need this molecular biology reagent tailored to your workflow? We can help with bulk or custom pack sizes, alternative formats (spin column, magnetic bead or reagent-only kits; premixed or stand-alone reference dye), matched extraction, cDNA synthesis and qPCR workflows, and sourcing of related enzymes, ladders and consumables not in our catalog. Click Talk to a Scientist to submit a request, email us at support@biohippo.com, or explore our Research Services — our team will follow up with feasibility details and next steps.
Sturgeon roe-derived tetrapeptide NLPL attenuates oxidative stress associated skin senescence endpoints, accompanied by increased NAD+, improved telomere-associated biomarkers, and enhanced ECM-related readouts. Journal of Future Foods, 2026. 10.1016/j.jfutfo.2026.03.012
Increased dairy product consumption is associated with shorter telomere length in buccal cells among normotensive adults. Biomedicines, 2026. 10.37796/2211-8039.1692
Effect of obesity and NAFLD on leukocyte telomere length and hTERT gene MNS16A VNTR variant. Scientific Reports, 2024. 10.1038/s41598-024-77091-5
Telomere Length and Mitochondrial Copy Number as Potential Biomarkers for Male Infertility in Iraqi Men. Genes, 2025. 10.3390/genes16121402